48T
$299.00
96T
$399.00
Species:
Human
Alternative Name:
MUC21, C6orf25, KMQK697, MUC-21, bCX31G15.2, mucin 21, cell surface associated
Uniport ID:
Q5SSG8
Assay Type:
Sandwich
unit:
ng/mL
Standard:
10ng/mL
Sensitivity:
0.1 ng/mL
Detection Range:
0.16-10ng/mL
Assay Time:
3.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=100μL
Research:
Signal transduction;Infection immunity;Hematology;Kidney biomarker;
Test Principle:
This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human MUC21. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human MUC21 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human MUC21, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human MUC21.You can calculate the concentration of Human MUC21 in the samples by comparing the OD of the samples to the standard curve.
| ng/mL | OD | Corrected |
|---|---|---|
| 10 | 2.396 | 2.358 |
| 5 | 1.711 | 1.673 |
| 2.5 | 0.995 | 0.957 |
| 1.25 | 0.488 | 0.45 |
| 0.63 | 0.249 | 0.211 |
| 0.32 | 0.126 | 0.088 |
| 0.16 | 0.079 | 0.041 |
| 0 | 0.038 | 0 |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean(ng/mL) | 0.18 | 2.33 | 4.66 | 0.18 | 1.77 | 4.85 |
| Standard deviation | 0.02 | 0.12 | 0.22 | 0.01 | 0.1 | 0.25 |
| C V (%) | 2.64 | 5.87 | 4.13 | 6.72 | 5.84 | 6.6 |
| Sample Type | Range(%) | Average Recovery(%) |
|---|---|---|
| Serum (n=8) | 80-95 | 86 |
| EDTA plasma (n=8) | 94-106 | 100 |
| Cell culture media (n=8) | 80-97 | 88 |
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 85-97 | 86-95 | 92-102 |
| Average (%) | 94 | 94 | 101 | |
| 1:4 | Range (%) | 87-96 | 93-103 | 81-93 |
| Average (%) | 92 | 100 | 90 | |
| 1:8 | Range (%) | 85-96 | 87-98 | 86-98 |
| Average (%) | 94 | 94 | 87 | |
| 1:16 | Range (%) | 85-95 | 85-96 | 91-103 |
| Average (%) | 93 | 93 | 96 |